This optical sensor measures the turbidity level in aqueous solutions. The wide measuring range enables its deployment in many applications. The measured value is automatically transferred via a digital interface to the connected JUMO AQUIS 500 RS indicating device with integrated controller.
general technical data | |
---|---|
Temperature sensor | Integrated NTC (Negative Temperature Coefficient) |
Application areas | • Municipal and industrial sewage treatment plants • Protection of water bodies • Fish farms • Process engineering plants |
Dimensions | Diameter: 27 mm, length: 170 mm |
Measuring error | < 5 % of the displayed measured value |
Voltage supply | DC 5 to 12 V |
Max. pressure | 5 bar |
Protection type | IP68 |
Material | PVC |
Operating temperature | 0 to 50 °C |
Measuring ranges | 4 measuring ranges: • 0 to 50 NTU • 0 to 200 NTU • 0 to 1000 NTU • 0 to 4000 NTU |
Interface | RS485 |
Features | Robust sensor, low-maintenance, calibration data and history in the sensor electronics, complete unit with JUMO AQUIS 500 RS (an indicating device with an integrated controller as well as 2 analog and 2 switching outputs) |
Resolution | 0.01 to 1 NTU (depending on the set measuring range) |
Measuring principle | Infrared measurement (880 nm) according to the 90° scattered-light principle (according to DIN EN ISO 7027) |
No related documents found.